What a skin-contact polyolefin is — and when to choose it
Polyolefins — polypropylene, polyethylene and their blends — are light, tough, fatigue-resistant and inert to sweat, soap and disinfectants. Choose a skin-contact grade over ordinary PP when the printed part will sit on a patient's or wearer's skin and your quality file needs material data. For device housings and lab parts that need stiffness rather than flexibility, medical-grade ABS is the neighbor.
“Skin-contact certified” usually means test specimens passed the three endpoints FDA and ISO 10993-1 associate with intact-skin surface devices: cytotoxicity (ISO 10993-5, in vitro), sensitization (ISO 10993-10) and irritation (ISO 10993-23, split out of -10 in 2021; older reports cite -10). FDA's ISO 10993-1 guidance goes further for intact skin: polypropylene and polyethylene are on its list of materials where a device maker may document materials instead of submitting test reports — although some cases, such as devices for neonates, call for discussion with FDA first.
None of this transfers to your part automatically. Results describe the specimen tested; printing parameters, colorants, cleaning and finishing are yours. The device manufacturer owns the biological evaluation of the finished device.
Typical properties
| Property | Typical range | Test method |
|---|---|---|
| Tensile strength (printed, XY) | 10–16 MPa | ISO 527 / ASTM D3039 |
| Elongation at break (XY) | Over 100% (grade-dependent) | ISO 527 |
| Heat deflection, 0.45 MPa | 54–65 °C | ISO 75 |
| Melting point | 130–165 °C | DSC |
| Density | 0.89–0.91 g/cm³ | ISO 1183 |
| Skin-contact data | Cytotoxicity, sensitization, irritation | ISO 10993-5 / -10 / -23 |
Typical published values across manufacturers, measured on test specimens; check your spool’s datasheet. Printed parts are usually weaker between layers (Z) than along them (XY).
Typical print settings
| Nozzle temperature | 200–240 °C |
|---|---|
| Build plate | 50–100 °C on PP sheet or PP tape |
| Chamber | Closed, warm; one maker recommends 65–80 °C (Epitum: up to 90 °C) |
| Print speed | 20–50 mm/s |
| Part cooling | 0–50% |
| Drying | 60–70 °C · 4–24 h if exposed to humidity |
| Nozzle | Clean, dedicated; ≥ 0.4 mm |
Starting point compiled from manufacturers’ recommendations — fine-tune with the Printoformer material profile for your spool, nozzle and part geometry.
Print a test coupon on your bed surface first: polyolefins bond only to polyolefins, so a PP sheet or PP tape is the foundation. Too hot a bed can weld the part to a PP sheet.
Which Epitum: JL or L
The Epitum L fits clinics and labs printing one material: its chamber, heated up to 90 °C, covers the warm build space polyolefins like, and 360 × 360 × 610 mm takes full-length leg orthoses. The Epitum JL keeps one nozzle dedicated to the skin-contact grade while the other runs other jobs. Expect moderate speeds, far below the 500 mm/s travel peak.
Applications
- Custom wrist, hand and finger splints
- Ankle-foot and other orthoses, from prototypes to fitted devices
- Cosmetic covers and padding shells for prostheses
- Wearable housings, straps and sports protective inserts
Troubleshooting and tips
- Warping on long parts. Close and heat the chamber, add a wide brim and avoid sharp corners.
- Rough edges. Smooth trim lines and strap slots; edges, not material, cause most skin complaints.
- Heat. Polyolefins deflect around 55–65 °C under light load, so steam sterilization and hot-car storage can distort parts.
- Keep it clean. No bed adhesives or coatings on skin-contact faces unless they are part of your evaluation.
Available from MG World Tech: Healthfil FORTIS, 0.75 kg spool, $93.54. Order or request a quote →
Frequently asked questions
What does skin-contact certified filament mean?
Test specimens of the filament passed cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10) and usually irritation (ISO 10993-23) tests. It describes the material, not your finished device.
Is polypropylene safe for skin contact?
PP has a long history in orthoses and is on FDA's list of low-risk materials for devices contacting only intact skin. Additives, colorants and processing still matter, so use a grade with test data.
Do I still need a biocompatibility evaluation for a printed orthosis?
Yes — the device maker evaluates the finished device under ISO 10993-1 and the applicable regulations. Material data and FDA's intact-skin approach can reduce, not remove, that work.
Why does polypropylene filament warp?
Polyolefins shrink as they crystallize. A closed, warm chamber, a PP bed surface and a brim keep parts flat; fiber-filled PP-CF warps much less but is too stiff for most skin-contact parts.